Gate timeline: 2026-10-29T1853Z_fw-b1435

Session2026-10-29T1853Z_fw-b1435
NotesTest firmware b1435 (g11d5c51): each step of a Vrect frame separated by 10 ms, MCU in Stop mode during the pauses. Three-firmware series, 5 min each, Gate down 5 min before each.
Gate log2026-10-29-18-59-56.txt
Gate1.1.0-dev.feature-implant-telemetry+b4443.g27b6a54
ClipPiece firmware1.1.0-dev+b1510.g5ba050d
Implant firmware1.1.0-dev.test-only-vrect-monitoring-pauses-10ms+b1435.g11d5c51
Implant FPGA3.0
Implant HW id0747373435333434002E0036
Wearable device id0447323532363232004A0036
Log2026-10-29 18:59:56.784 – 19:06:13.367 UTC
Acquisition19:00:06.493 – 19:06:13.367 (367 s, log end)
Recording window19:00:14.320 – 19:05:13.660 (299.3 s, session.json)
State at log endphase acquiring, health operational, acquisition acquiring

Events

Gate start and restarts, workflow-state changes (gate-core), non-INFO entries, data-quality reports, gate-core protocol issues, ClipPiece MCU firmware states that change. Telemetry is in the plots.

UTCKindWhatMessage
18:59:57.764GATE_VERSIONGate start; phase → starting; health → starting
Gate version: 1.1.0
Gate version: 1.1.0
18:59:56.784CONNECTION_SUCCESSdata_link → connected
Connected to Data Port
Connected to Data Port
18:59:57.764GATE_COMPATIBLEphase → waiting_for_wearable; health → operational; acquisition → stopped
Gate (version = 1.1.0) is compatible
Gate (version = 1.1.0) is compatible
18:59:57.775CONNECTION_SUCCESSphase → wearable_startup; control_link → connected; wearable → connected
Successfully established connection with the USB Configuration Link (USB device /dev/ttyUS
Successfully established connection with the USB Configuration Link (USB device /dev/ttyUSB0)
18:59:58.276COMMAND_TIMEOUTERROR
Command [MID=0x41][RID=0x01] has timeout
Command [MID=0x41][RID=0x01] has timeout
18:59:58.306WEARABLE_COMPATIBLEwearable → compatible
Gate (version = 1.1.0) is compatible with the Wearable
Gate (version = 1.1.0) is compatible with the Wearable
19:00:00.610IMPLANT_INFOphase → implant_startup; implant → connected
Implant start-up checks were completed in less than 10 seconds:
Implant start-up checks were completed in less than 10 seconds:
IMP_FW_STATE_STARTUP_CHECKS_DONE = completed
IMP_FW_STATE_BANK1_INVALID = valid
IMP_FW_STATE_BANK2_INVALID = valid
IMP_FW_STATE_RUNNING_BANK = Bank 1
IMP_FW_STATE_BANKS_DIFF = identical
IMP_FW_STATE_NOT_COMP_WITH_HW = compatible
IMP_FPGA_STATE_STARTUP_CHECKS_DONE = completed
IMP_FPGA_STATE_BITSTREAM_INVALID = valid
IMP_SENSORS_STATE_STARTUP_CHECKS_DONE = completed
IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE = implausible
19:00:01.111IMPLANT_COMPATIBLEimplant → compatible
Gate (version = 1.1.0) is compatible with both the Wearable and the Implant
Gate (version = 1.1.0) is compatible with both the Wearable and the Implant
19:00:01.887WEARABLE_INFOphase → optimizing; implant → optimizing; acquisition → starting
ClipPiece MCU Firmware States: Implant states:
ClipPiece MCU Firmware States: Implant states:
- IMP_FW_STATE_STARTUP_CHECKS_DONE: in progress
- IMP_FW_STATE_BANK1_INVALID: valid
- IMP_FW_STATE_BANK2_INVALID: valid
- IMP_FW_STATE_RUNNING_BANK: Bank 1
- IMP_FW_STATE_BANKS_DIFF: identical
- IMP_FW_STATE_NOT_COMP_WITH_HW: compatible
- IMP_FPGA_STATE_STARTUP_CHECKS_DONE: not completed
- IMP_FPGA_STATE_BITSTREAM_INVALID: valid
- IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: in progress
- IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: plausible

Wearable states:
- STARTUP_CHECKS_DONE_CP_STATE: completed
- BANK1_INVALID_CP_STATE: valid
- BANK2_INVALID_CP_STATE: valid
- RUN_BANK_CP_STATE: Bank 1
- BANKS_CRC_DIFF_CP_STATE: identical
- FW_NOT_COMP_STATE: compatible
- STARTUP_CHECKS_TIMEOUT_CP_STATE: allocated time not reached
- WPT_INVALID_CP_STATE: in allowed range
- RFON_INVALID_CP_STATE: valid
- ENABLE_RFON_INVALID_CP_STATE: valid
- WPT_PARAM_CHECK_STATE: check passed
19:00:05.895OPTIMIZATION_SUCCESSimplant → connected
Optical Link Optimization was a success. The following combination of parameters result in
Optical Link Optimization was a success. The following combination of parameters result in a current consumption of 1016µA and RSSI of 547:
- VCSEL 0: enabled
- VCSEL 1: disabled
- VCSEL 2: disabled
- VCSEL 3: disabled
- power level: 75
19:00:06.493ACQUIRINGphase → acquiring; acquisition → acquiring
Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=1
Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
19:00:06.583MCU_FW_STATESIMP_FW_STATE_STARTUP_CHECKS_DONE: in progress → completed; IMP_FPGA_STATE_STARTUP_CHECKS_DONE: not completed → completed; IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: in progress → completed; IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: plausible → implausible
Implant states: IMP_FW_STATE_STARTUP_CHECKS_DONE: completed, IMP_FW_STATE_BANK1_INVALID: v
Implant states: IMP_FW_STATE_STARTUP_CHECKS_DONE: completed, IMP_FW_STATE_BANK1_INVALID: valid, IMP_FW_STATE_BANK2_INVALID: valid, IMP_FW_STATE_RUNNING_BANK: Bank 1, IMP_FW_STATE_BANKS_DIFF: identical, IMP_FW_STATE_NOT_COMP_WITH_HW: compatible, IMP_FPGA_STATE_STARTUP_CHECKS_DONE: completed, IMP_FPGA_STATE_BITSTREAM_INVALID: valid, IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: completed, IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: implausible; Wearable states: STARTUP_CHECKS_DONE_CP_STATE: completed, BANK1_INVALID_CP_STATE: valid, BANK2_INVALID_CP_STATE: valid, RUN_BANK_CP_STATE: Bank 1, BANKS_CRC_DIFF_CP_STATE: identical, FW_NOT_COMP_STATE: compatible, STARTUP_CHECKS_TIMEOUT_CP_STATE: allocated time not reached, WPT_INVALID_CP_STATE: in allowed range, RFON_INVALID_CP_STATE: valid, ENABLE_RFON_INVALID_CP_STATE: valid, WPT_PARAM_CHECK_STATE: check passed

Over time

Gate phases and telemetry over time