2026-10-30T1001Z_eco ==================== gate 1.1.0-dev.feature-implant-telemetry+b4443.g27b6a54 clippiece_fw 1.1.0-dev+b1510.g5ba050d implant_fw 1.1.0-dev.ablapp-1502-keepalive-7s+b1413.gb44b676 implant_fpga 3.0 implant_hw_id 04473235323632320045002E wearable_device_id 0447323532363232004A0036 implant_fpga_crc32 0xA50ED517 recorded 299.339 s from 2026-10-30T10:10:33.700991+00:00, 598684 samples, 2747 notifications recorded: 598684 samples over 299.339 s, 2747 notifications; analysing after the first 10 s NEURAL TIMESTAMPS (as stored) samples 578683, span 289.3391 s, rate from first-last 2000.0131 Hz, fitted 2000.0126 Hz step: min 0.4999 ms, median 0.5000 ms, max 0.5001 ms; steps != 0.5 ms (±1 µs): 0; non-increasing: 0; gaps > 1 ms: 0 residual vs straight line: p1 -0.0678 ms, p99 +0.0456 ms, max |0.1251| ms status channel: {0.0: 578683} NOTIFICATIONS 2655 after skip; out of order (stamp earlier than previous): 0 Vrect: 577, MCU_INPUT_OUTPUT: 577, MCU_FW_STATES: 577, Main current: 289, Transmitter current: 289, Safety state: 289, Temperature: 57 SPIKES: 0 on >= 100 channels notification n with spike spike peak -> notification, ms: p5 / median / p95 Vrect 577 0 - Main current 289 0 - Transmitter current 289 0 - Safety state 289 0 - Temperature 57 0 - 7 s CANDIDATES (no notification 2-9 ms after, z > 40): [] s after analysis start AMPLITUDES (median channel, µV, median over spikes) background: 4235.8 µV per channel GATE LOG: acquisitions, data losses, errors (whole log of this Gate run) [2026-10-30T10:10:20.508Z] [ERROR] [COMMAND_TIMEOUT] Command [MID=0x41][RID=0x01] has timeout [2026-10-30T10:10:27.872Z] [INFO] [ACQUIRING] Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFF