Gate timeline: 2026-10-30T1556Z_baseline

Session2026-10-30T1556Z_baseline
NotesSaline bath, 30 Oct afternoon; series b1441, b1436, b1413 (5 min down, 5 min recorded)
Gate log2026-10-30-16-02-35.txt
Gate1.1.0-dev.feature-implant-telemetry+b4443.g27b6a54
ClipPiece firmware1.1.0-dev+b1510.g5ba050d
Implant firmware1.1.0-dev.ablapp-1502-keepalive-7s+b1413.gb44b676
Implant FPGA3.0
Implant HW id0747373435333434002E0036
Wearable device id0447323532363232004A0036
Log2026-10-30 16:02:35.124 – 16:09:00.827 UTC
Acquisition16:02:43.695 – 16:09:00.827 (377 s, log end)
Recording window16:02:50.385 – 16:07:49.724 (299.3 s, session.json)
State at log endphase acquiring, health operational, acquisition acquiring

Events

Gate start and restarts, workflow-state changes (gate-core), non-INFO entries, data-quality reports, gate-core protocol issues, ClipPiece MCU firmware states that change. Telemetry is in the plots.

UTCKindWhatMessage
16:02:36.104GATE_VERSIONGate start; phase → starting; health → starting
Gate version: 1.1.0
Gate version: 1.1.0
16:02:35.124CONNECTION_SUCCESSdata_link → connected
Connected to Data Port
Connected to Data Port
16:02:36.104GATE_COMPATIBLEphase → waiting_for_wearable; health → operational; acquisition → stopped
Gate (version = 1.1.0) is compatible
Gate (version = 1.1.0) is compatible
16:02:36.116CONNECTION_SUCCESSphase → wearable_startup; control_link → connected; wearable → connected
Successfully established connection with the USB Configuration Link (USB device /dev/ttyUS
Successfully established connection with the USB Configuration Link (USB device /dev/ttyUSB0)
16:02:36.617COMMAND_TIMEOUTERROR
Command [MID=0x41][RID=0x01] has timeout
Command [MID=0x41][RID=0x01] has timeout
16:02:36.647WEARABLE_COMPATIBLEwearable → compatible
Gate (version = 1.1.0) is compatible with the Wearable
Gate (version = 1.1.0) is compatible with the Wearable
16:02:38.951IMPLANT_INFOphase → implant_startup; implant → connected
Implant start-up checks were completed in less than 10 seconds:
Implant start-up checks were completed in less than 10 seconds:
IMP_FW_STATE_STARTUP_CHECKS_DONE = completed
IMP_FW_STATE_BANK1_INVALID = valid
IMP_FW_STATE_BANK2_INVALID = valid
IMP_FW_STATE_RUNNING_BANK = Bank 1
IMP_FW_STATE_BANKS_DIFF = identical
IMP_FW_STATE_NOT_COMP_WITH_HW = compatible
IMP_FPGA_STATE_STARTUP_CHECKS_DONE = completed
IMP_FPGA_STATE_BITSTREAM_INVALID = valid
IMP_SENSORS_STATE_STARTUP_CHECKS_DONE = completed
IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE = implausible
16:02:39.451IMPLANT_COMPATIBLEimplant → compatible
Gate (version = 1.1.0) is compatible with both the Wearable and the Implant
Gate (version = 1.1.0) is compatible with both the Wearable and the Implant
16:02:40.195WEARABLE_INFOphase → optimizing; implant → optimizing; acquisition → starting
ClipPiece MCU Firmware States: Implant states:
ClipPiece MCU Firmware States: Implant states:
- IMP_FW_STATE_STARTUP_CHECKS_DONE: in progress
- IMP_FW_STATE_BANK1_INVALID: valid
- IMP_FW_STATE_BANK2_INVALID: valid
- IMP_FW_STATE_RUNNING_BANK: Bank 1
- IMP_FW_STATE_BANKS_DIFF: identical
- IMP_FW_STATE_NOT_COMP_WITH_HW: compatible
- IMP_FPGA_STATE_STARTUP_CHECKS_DONE: not completed
- IMP_FPGA_STATE_BITSTREAM_INVALID: valid
- IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: in progress
- IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: plausible

Wearable states:
- STARTUP_CHECKS_DONE_CP_STATE: completed
- BANK1_INVALID_CP_STATE: valid
- BANK2_INVALID_CP_STATE: valid
- RUN_BANK_CP_STATE: Bank 1
- BANKS_CRC_DIFF_CP_STATE: identical
- FW_NOT_COMP_STATE: compatible
- STARTUP_CHECKS_TIMEOUT_CP_STATE: allocated time not reached
- WPT_INVALID_CP_STATE: in allowed range
- RFON_INVALID_CP_STATE: valid
- ENABLE_RFON_INVALID_CP_STATE: valid
- WPT_PARAM_CHECK_STATE: check passed
16:02:43.169OPTIMIZATION_SUCCESSimplant → connected
Optical Link Optimization was a success. The following combination of parameters result in
Optical Link Optimization was a success. The following combination of parameters result in a current consumption of 950µA and RSSI of 567:
- VCSEL 0: enabled
- VCSEL 1: disabled
- VCSEL 2: disabled
- VCSEL 3: disabled
- power level: 72
16:02:43.695ACQUIRINGphase → acquiring; acquisition → acquiring
Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=1
Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
16:02:43.901MCU_FW_STATESIMP_FW_STATE_STARTUP_CHECKS_DONE: in progress → completed; IMP_FPGA_STATE_STARTUP_CHECKS_DONE: not completed → completed; IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: in progress → completed; IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: plausible → implausible
Implant states: IMP_FW_STATE_STARTUP_CHECKS_DONE: completed, IMP_FW_STATE_BANK1_INVALID: v
Implant states: IMP_FW_STATE_STARTUP_CHECKS_DONE: completed, IMP_FW_STATE_BANK1_INVALID: valid, IMP_FW_STATE_BANK2_INVALID: valid, IMP_FW_STATE_RUNNING_BANK: Bank 1, IMP_FW_STATE_BANKS_DIFF: identical, IMP_FW_STATE_NOT_COMP_WITH_HW: compatible, IMP_FPGA_STATE_STARTUP_CHECKS_DONE: completed, IMP_FPGA_STATE_BITSTREAM_INVALID: valid, IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: completed, IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: implausible; Wearable states: STARTUP_CHECKS_DONE_CP_STATE: completed, BANK1_INVALID_CP_STATE: valid, BANK2_INVALID_CP_STATE: valid, RUN_BANK_CP_STATE: Bank 1, BANKS_CRC_DIFF_CP_STATE: identical, FW_NOT_COMP_STATE: compatible, STARTUP_CHECKS_TIMEOUT_CP_STATE: allocated time not reached, WPT_INVALID_CP_STATE: in allowed range, RFON_INVALID_CP_STATE: valid, ENABLE_RFON_INVALID_CP_STATE: valid, WPT_PARAM_CHECK_STATE: check passed

Over time