2026-10-30T1556Z_baseline ========================= gate 1.1.0-dev.feature-implant-telemetry+b4443.g27b6a54 clippiece_fw 1.1.0-dev+b1510.g5ba050d implant_fw 1.1.0-dev.ablapp-1502-keepalive-7s+b1413.gb44b676 implant_fpga 3.0 implant_hw_id 0747373435333434002E0036 wearable_device_id 0447323532363232004A0036 implant_fpga_crc32 0xA50ED517 recorded 299.339 s from 2026-10-30T16:02:50.385653+00:00, 598860 samples, 2747 notifications recorded: 598860 samples over 299.339 s, 2747 notifications; analysing after the first 10 s RECORDING samples 578854 over 289.339 s, 2000.6030 samples/s on the recording clock steps: median 0.4998 ms; not 0.5 ms (±1 µs): 0; non-increasing: 0; gaps > 1 ms: 0; jitter p1/p99 -0.051 / +0.021 ms status channel: {0.0: 578854} notifications: 2655, out of order: 0; IMPLANT_RECTIFIED_VOLTAGE: 577, MCU_INPUT_OUTPUT: 577, MCU_FW_STATES: 577, IMPLANT_MAIN_CURRENT: 289, IMPLANT_TRANSMITTER_CURRENT: 289, IMPLANT_SAFETY_STATE: 289, IMPLANT_TEMPERATURE: 57 MEASUREMENTS (µV; median channel of each group) All Intan 1 Intan 2 Vrect message spike 52.86 59.17 53.47 1 s burst spike 129.82 133.80 123.16 Temperature message spike 75.40 83.07 74.26 7 s keep-alive 210.43 212.29 186.73 976 Hz line 1.78 1.98 2.04 122 Hz line 0.36 0.40 0.39 Noise, per channel 3.06 3.12 2.98 Noise, median channel 2.18 2.53 2.54 Mains, 50 Hz (RMS) 0.09 0.14 0.11 * below its floor: not distinguishable from noise (measures.py) clipped samples 0.0%; averages over 512 Vrect, 194 burst, 30 temperature messages (alone) keep-alives found: 41 (about 41 expected), intervals 6.998-7.002 s GATE LOG: acquisitions, data losses, errors (whole log of this Gate run) [2026-10-30T16:02:36.617Z] [ERROR] [COMMAND_TIMEOUT] Command [MID=0x41][RID=0x01] has timeout [2026-10-30T16:02:43.695Z] [INFO] [ACQUIRING] Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFF