| Session | 2026-11-01T1842Z_always-awake |
|---|
| Notes | 1 Nov evening series on Gate b4449 (caac85f, FPGA identity read before the FPGA version): b1446, b1441, b1436, b1413, each flashed, 5 min Gate stopped, 5 min recorded |
|---|
| Gate log | 2026-11-01-18-49-10.txt |
|---|
| Gate | 1.1.0-dev.feature-implant-telemetry+b4449.gcaac85f |
|---|
| ClipPiece firmware | 1.1.0-dev+b1510.g5ba050d |
|---|
| Implant firmware | 1.1.0-dev.test-only-no-mcu-low-power+b1436.gf7e8df3 |
|---|
| Implant FPGA | 3.0 |
|---|
| Implant FPGA bitstream | 135180 bytes, CRC32 0xA50ED517 |
|---|
| Implant HW id | 0747373435333434002E0036 |
|---|
| Wearable device id | 0447323532363232004A0036 |
|---|
| Log | 2026-11-01 18:49:10.063 – 18:55:25.579 UTC |
|---|
| Acquisition | 18:49:18.541 – 18:55:25.579 (367 s, log end) |
|---|
| Recording window | 18:49:24.753 – 18:54:24.092 (299.3 s, session.json) |
|---|
| State at log end | phase acquiring, health operational, acquisition acquiring |
|---|
Gate start and restarts, workflow-state changes (gate-core), non-INFO entries, data-quality reports,
gate-core protocol issues, ClipPiece MCU firmware states that change. Telemetry is in the plots.
| UTC | Kind | What | Message |
|---|
| 18:49:11.042 | GATE_VERSION | Gate start; phase → starting; health → starting | Gate version: 1.1.0Gate version: 1.1.0 |
| 18:49:10.063 | CONNECTION_SUCCESS | data_link → connected | Connected to Data PortConnected to Data Port |
| 18:49:11.043 | GATE_COMPATIBLE | phase → waiting_for_wearable; health → operational; acquisition → stopped | Gate (version = 1.1.0) is compatibleGate (version = 1.1.0) is compatible |
| 18:49:11.054 | CONNECTION_SUCCESS | phase → wearable_startup; control_link → connected; wearable → connected | Successfully established connection with the USB Configuration Link (USB device /dev/ttyUSSuccessfully established connection with the USB Configuration Link (USB device /dev/ttyUSB0) |
| 18:49:11.555 | COMMAND_TIMEOUT | ERROR | Command [MID=0x41][RID=0x01] has timeoutCommand [MID=0x41][RID=0x01] has timeout |
| 18:49:11.585 | WEARABLE_COMPATIBLE | wearable → compatible | Gate (version = 1.1.0) is compatible with the WearableGate (version = 1.1.0) is compatible with the Wearable |
| 18:49:13.890 | IMPLANT_INFO | phase → implant_startup; implant → connected | Implant start-up checks were completed in less than 10 seconds:Implant start-up checks were completed in less than 10 seconds:
IMP_FW_STATE_STARTUP_CHECKS_DONE = completed
IMP_FW_STATE_BANK1_INVALID = valid
IMP_FW_STATE_BANK2_INVALID = valid
IMP_FW_STATE_RUNNING_BANK = Bank 1
IMP_FW_STATE_BANKS_DIFF = identical
IMP_FW_STATE_NOT_COMP_WITH_HW = compatible
IMP_FPGA_STATE_STARTUP_CHECKS_DONE = completed
IMP_FPGA_STATE_BITSTREAM_INVALID = valid
IMP_SENSORS_STATE_STARTUP_CHECKS_DONE = completed
IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE = implausible |
| 18:49:14.494 | IMPLANT_COMPATIBLE | implant → compatible | Gate (version = 1.1.0) is compatible with both the Wearable and the ImplantGate (version = 1.1.0) is compatible with both the Wearable and the Implant |
| 18:49:15.237 | WEARABLE_INFO | phase → optimizing; implant → optimizing; acquisition → starting | ClipPiece MCU Firmware States: Implant states:ClipPiece MCU Firmware States: Implant states:
- IMP_FW_STATE_STARTUP_CHECKS_DONE: in progress
- IMP_FW_STATE_BANK1_INVALID: valid
- IMP_FW_STATE_BANK2_INVALID: valid
- IMP_FW_STATE_RUNNING_BANK: Bank 1
- IMP_FW_STATE_BANKS_DIFF: identical
- IMP_FW_STATE_NOT_COMP_WITH_HW: compatible
- IMP_FPGA_STATE_STARTUP_CHECKS_DONE: not completed
- IMP_FPGA_STATE_BITSTREAM_INVALID: valid
- IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: in progress
- IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: plausible
Wearable states:
- STARTUP_CHECKS_DONE_CP_STATE: completed
- BANK1_INVALID_CP_STATE: valid
- BANK2_INVALID_CP_STATE: valid
- RUN_BANK_CP_STATE: Bank 1
- BANKS_CRC_DIFF_CP_STATE: identical
- FW_NOT_COMP_STATE: compatible
- STARTUP_CHECKS_TIMEOUT_CP_STATE: allocated time not reached
- WPT_INVALID_CP_STATE: in allowed range
- RFON_INVALID_CP_STATE: valid
- ENABLE_RFON_INVALID_CP_STATE: valid
- WPT_PARAM_CHECK_STATE: check passed
|
| 18:49:18.005 | OPTIMIZATION_SUCCESS | implant → connected | Optical Link Optimization was a success. The following combination of parameters result inOptical Link Optimization was a success. The following combination of parameters result in a current consumption of 966µA and RSSI of 564:
- VCSEL 0: enabled
- VCSEL 1: disabled
- VCSEL 2: disabled
- VCSEL 3: disabled
- power level: 74 |
| 18:49:18.541 | ACQUIRING | phase → acquiring; acquisition → acquiring | Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=1Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF |
| 18:49:18.748 | MCU_FW_STATES | IMP_FW_STATE_STARTUP_CHECKS_DONE: in progress → completed; IMP_FPGA_STATE_STARTUP_CHECKS_DONE: not completed → completed; IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: in progress → completed; IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: plausible → implausible | Implant states: IMP_FW_STATE_STARTUP_CHECKS_DONE: completed, IMP_FW_STATE_BANK1_INVALID: vImplant states: IMP_FW_STATE_STARTUP_CHECKS_DONE: completed, IMP_FW_STATE_BANK1_INVALID: valid, IMP_FW_STATE_BANK2_INVALID: valid, IMP_FW_STATE_RUNNING_BANK: Bank 1, IMP_FW_STATE_BANKS_DIFF: identical, IMP_FW_STATE_NOT_COMP_WITH_HW: compatible, IMP_FPGA_STATE_STARTUP_CHECKS_DONE: completed, IMP_FPGA_STATE_BITSTREAM_INVALID: valid, IMP_SENSORS_STATE_STARTUP_CHECKS_DONE: completed, IMP_SENSORS_STATE_IMPLAUSIBLE_TEMPERATURE: implausible; Wearable states: STARTUP_CHECKS_DONE_CP_STATE: completed, BANK1_INVALID_CP_STATE: valid, BANK2_INVALID_CP_STATE: valid, RUN_BANK_CP_STATE: Bank 1, BANKS_CRC_DIFF_CP_STATE: identical, FW_NOT_COMP_STATE: compatible, STARTUP_CHECKS_TIMEOUT_CP_STATE: allocated time not reached, WPT_INVALID_CP_STATE: in allowed range, RFON_INVALID_CP_STATE: valid, ENABLE_RFON_INVALID_CP_STATE: valid, WPT_PARAM_CHECK_STATE: check passed |