 "_about": "Implant MCU firmwares, by build number (the +bNNNN in the build identity). name: how pages and plots call it; idle: what the MCU does between tasks; colour: index into palette.json, fixed per firmware so that a firmware looks the same in every view; default: selected when the dashboard opens.",
 "b1413": {"name": "Baseline", "idle": "Low-power idle", "description": "The MCU enters Stop or Sleep mode between tasks.", "nickname": "eco", "colour": 0, "default": true},
 "b1436": {"name": "Always-awake", "idle": "Running idle", "description": "Baseline with the MCU kept running between tasks; it never enters Stop or Sleep mode. Test build.", "nickname": "turbo", "colour": 1, "default": true},
 "b1435": {"name": "Paced Vrect", "idle": "Low-power idle", "description": "Vrect message sent in steps 10 ms apart; the MCU enters Stop mode in the pauses. Test build.", "colour": 2},
 "b1437": {"name": "Paced Vrect, awake", "idle": "Running idle", "description": "Vrect message sent in steps 10 ms apart; the MCU keeps running in the pauses. Test build.", "colour": 3},
 "b1439": {"name": "Light-sleep", "idle": "Sleep-mode idle", "description": "The MCU enters Sleep mode instead of Stop mode between its 1 ms wake-ups: the main regulator and the 16 MHz clock stay on, only the CPU clock stops. Test build.", "colour": 4, "default": true},
 "b1441": {"name": "Slow-awake", "idle": "Running idle at 2 MHz", "description": "During the acquisition the MCU never enters a low-power mode and its CPU runs at a constant 2 MHz (16 MHz / 8), sensor frames included (their SPI at 125 kHz, about 8 times longer); HKL requests at 16 MHz; outside the acquisition as develop. Test build.", "colour": 5, "default": true},
 "b1446": {"name": "Slow-awake, no 1 ms tick", "idle": "Running idle at 2 MHz, no 1 ms wake-up", "description": "Slow-awake (b1441) without the 1024 Hz wake-up during the acquisition: between two broadcast events the CPU stays in one busy wait at 2 MHz, ended by a single LPTIM match at the next sensor frame or keep-alive, the 1 Hz counter wrap or an HKL byte; the safety comparators are polled at this lower rate. Outside the acquisition as develop. Test build.", "colour": 6, "default": true},
 "b1457": {
  "name": "Low-noise acquisition",
  "idle": "Slow clock, no 1 ms wake-up during acquisition",
  "description": "ABLAPP-1511 candidate for saline tests. During acquisition the MCU runs at a slow clock and wakes for the next sensor broadcast instead of every 1 ms; HKL requests are still answered at full speed.",
  "colour": 7,
  "default": true,
  "build_identity": "1.1.0-dev.ablapp-1511-implant-low-noise-acquisition+b1457.gbdd6c81",
  "validation_crc32": "0x8C3E0BD1",
  "size_bytes": 22076,
  "sha256": "0cded7c6d8cff010ed7b85f3e5f23b9b30c3a0eb3bfcf4d198ff35ea11588ef1",
  "pack": "implant-fw-dev-pack.b1457.ABLAPP-1511-implant-low-noise-acquisition.zip",
  "reported_3d_bench_result": "Firmware release report: the Vrect 1024 Hz line drops from about 5.7 mV to 0.045 mV (about 125 times), with the noise floor unchanged. This is a separate 3D-bench observation, not a saline-bath measurement."
 }
}