2026-11-04T1106Z_fw-b1436 ========================= gate 1.1.0-dev.feature-implant-telemetry+b4449.gcaac85f clippiece_fw 1.1.0-dev+b1510.g5ba050d implant_fw 1.1.0-dev.test-only-no-mcu-low-power+b1436.gf7e8df3 implant_fpga 3.0 implant_hw_id 0747373435333434002E0036 wearable_device_id 0447323532363232004A0036 implant_fpga_size_bytes 135180 implant_fpga_crc32 0xA50ED517 recorded 299.341 s from 2026-11-04T11:12:51.775285+00:00, 598866 samples, 2748 notifications recorded: 598866 samples over 299.341 s, 2748 notifications; analysing after the first 10 s RECORDING samples 578859 over 289.341 s, 2000.6107 samples/s on the recording clock steps: median 0.4998 ms; not 0.5 ms (±1 µs): 0; non-increasing: 0; gaps > 1 ms: 0; jitter p1/p99 -0.073 / +0.027 ms status channel: {0.0: 578859} notifications: 2656, out of order: 0; IMPLANT_RECTIFIED_VOLTAGE: 577, MCU_INPUT_OUTPUT: 577, MCU_FW_STATES: 577, IMPLANT_MAIN_CURRENT: 289, IMPLANT_TRANSMITTER_CURRENT: 289, IMPLANT_SAFETY_STATE: 289, IMPLANT_TEMPERATURE: 58 MEASUREMENTS (µV; median channel of each group) All Intan 1 Intan 2 Vrect message peak 4.77 5.27 4.58 1 s burst peak 11.43 12.67 10.72 Temperature message peak 8.03 9.06 7.36 7 s keep-alive 472.92 479.50 405.07 976 Hz line 0.33 0.36 0.33 122 Hz line 1.23 1.38 1.17 Noise, per channel 3.02 3.05 2.97 Noise, median channel 1.98 2.40 2.17 Mains, 50 Hz (RMS) 0.12 0.15 0.11 * within ± its floor: not distinguishable from noise (measures.py) clipped samples 0.0%; averages over 482 Vrect, 233 burst, 34 temperature messages (alone) keep-alives found: 41 (about 41 expected), intervals 6.998-7.002 s GATE LOG: acquisitions, data losses, errors (whole log of this Gate run) [2026-11-04T11:12:36.016Z] [ERROR] [COMMAND_TIMEOUT] Command [MID=0x41][RID=0x01] has timeout [2026-11-04T11:12:43.420Z] [INFO] [ACQUIRING] Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFF