2026-11-04T1255Z_fw-b1457 ========================= gate 1.1.0-dev.feature-implant-telemetry+b4449.gcaac85f clippiece_fw 1.1.0-dev+b1510.g5ba050d implant_fw 1.1.0-dev.ablapp-1511-implant-low-noise-acquisition+b1457.gbdd6c81 implant_fpga 3.0 implant_hw_id 0747373435333434002E0036 wearable_device_id 0447323532363232004A0036 implant_fpga_size_bytes 135180 implant_fpga_crc32 0xA50ED517 recorded 299.342 s from 2026-11-04T13:02:35.730950+00:00, 598867 samples, 2725 notifications recorded: 598867 samples over 299.342 s, 2725 notifications; analysing after the first 10 s RECORDING samples 578860 over 289.341 s, 2000.6108 samples/s on the recording clock steps: median 0.4998 ms; not 0.5 ms (±1 µs): 0; non-increasing: 0; gaps > 1 ms: 0; jitter p1/p99 -0.071 / +0.032 ms status channel: {0.0: 578860} notifications: 2633, out of order: 1; IMPLANT_RECTIFIED_VOLTAGE: 572, MCU_INPUT_OUTPUT: 572, MCU_FW_STATES: 572, IMPLANT_MAIN_CURRENT: 287, IMPLANT_TRANSMITTER_CURRENT: 286, IMPLANT_SAFETY_STATE: 286, IMPLANT_TEMPERATURE: 58 MEASUREMENTS (µV; median channel of each group) All Intan 1 Intan 2 Vrect message peak -2.12 -2.59 -2.12 1 s burst peak 3.03 3.24 3.10 Temperature message peak 3.38 3.69 3.33 7 s keep-alive 449.85 458.68 385.85 976 Hz line 0.18 0.21 0.19 122 Hz line 1.23 1.38 1.17 Noise, per channel 3.00 3.02 2.94 Noise, median channel 1.92 2.35 2.15 Mains, 50 Hz (RMS) 0.12 0.13 0.11 * within ± its floor: not distinguishable from noise (measures.py) clipped samples 0.0%; averages over 196 Vrect, 62 burst, 33 temperature messages (alone) keep-alives found: 41 (about 41 expected), intervals 7.000-7.004 s GATE LOG: acquisitions, data losses, errors (whole log of this Gate run) [2026-11-04T13:02:20.345Z] [ERROR] [COMMAND_TIMEOUT] Command [MID=0x41][RID=0x01] has timeout [2026-11-04T13:02:27.610Z] [INFO] [ACQUIRING] Neural data acquisition started with the Implant Configuration SR=2, HPF=24, LPF=11, DSP=16, electrode_list=0xFFFF